A Study on Measurement of the Half Value Layer in Diagnostic X-ray

진단용(診斷用) X선(線)의 반가층(半價層)에 관(關)한 연구(硏究)

  • Ko, Shin-Gwan (Dept. of Radiotechnology, Dae Jon Junior College)
  • 고신관 (대전보건전문대학 방사선과)
  • Published : 1984.12.15

Abstract

The quality of continuous x-ray beam depends upon the half value layer which varies according to the geometric conditions, the filtration thickness, and the amount of accelerated voltage (KVP). Experiments were conducted on the amount of electric energy that was changed to x-ray energy, and on the relationship between KVp and the intensity of x-rays. The results were as follows: 1. The amount of x-rays were not equal under the condition of the same exposure factor. 2. The intensity of x-rays was attenuated by an exponential function the geometric conditions were "good" and it was not when they were "poor". 3. The thicker the total filtration substance was and the higher the KVp was, the bigger the amount of x-ray energy was. 4. The homogeneity of medium energy x-ray was the best, when the total filtration substance was 3.9mm A1. 5. The mean energy of continuous x-ray was about 45% of KVp.