Characteristics and Applications of Soild State Nuclear Track Detectors -The Detection and Dosimetry of N-Ions by CR-39-

고체비적검출기(固體飛跡檢出器)의 특성(特性)과 그 응용(應用) -CR-39에 의한 질소(窒素)이온 검출(檢出)과 선양측정(線量測定)-

  • Published : 1984.12.15

Abstract

The optimum etching condition of allyl diglycol carbonate (CR-39) for detecting the 60 MeV N-ions was determined as $70^{\circ}C$, 20% NaOH for 130min, by considering the variations of track density and diameter. Under these conditions, the maximum detectable track density was $1.7{\times}10^7tr/cm^2$. Track densities were linearly increased with increase of the total charge of the incident 60 MeV N-ions. By considering the scattring of N-ions as the Rutherford elastic scattering of point source, the measured relative probability was well consistent with the calculated value. The detection efficiency of CR-39 was in the range of 54-41% for 60 MeV N-ions of 12 nC-100 nC.

Allyl diglycol carbonate(CR-39)에 조사(照射)된 60MeV의 질소이온을 검출하기 위한 최적부식 조건을 부식시간에 대한 비적밀도와 비적직경의 변화를 고려하여 $70^{\circ}C$, 20% NaOH 용액에 대해 부식시간 130분으로 정하였다. 이 조건에서 검출가능한 최대비적밀도는 $1.7{\times}10^7tr/cm^2$였으며 입사(入射)된 질소 이온의 총전하량의 증가에 따라 비적밀도가 선형적으로 증가하였다. 또한 $2{\mu}m$의 금박에 의한 질소 이온의 산란을 점선원(點線源)에 의한 Rutherford 산란으로 고려함으로써 측정된 상대산란확률이 이론치와 일치하여 선량측정이 가능하였으며 $12nC{\sim}100nC$의 입사(入射) 질소이온의 전하에 대해 $54{\sim}41%$의 검출 효율을 얻었다.