Ventricular pacing with screw-in sutureless myocardial electrode

무봉합심근전극을 사용한 인공심박조정기 설치40례 보고

  • Published : 1983.03.01

Abstract

We implanted pacemakers using screw-in sutureless myocardial electrode in 40 patients between February, 1980 and November, 1982 at Seoul Nation University Hospital. Mean age of total patient who received the pacemaker implantation was 39.9 years, but excluding the patient who received the open heart surgery, mean age was 49.3 years. Cause of pacemaker implantation was complete A-V block in 13 patients, symptomatic sick sinus syndrome in 12 patients, complete A-V block after open heart surgery in 12 patients, second degree A-V block with dizziness in one patient, and atrial fibrillation with slow ventricular response in two patients. In thirty nine cases, electrodes were implanted by subxiphoid approach, on the diaphragmatic surface of right ventricle. I n one case, electrode was implanted through the left anterior thoracotomy. Acute stimulation threshold, measured in 35 patients, varied from 0.1 to 2.5 mA [mean 0.85 mA.], and R-wave amplitude [sensitivity], measured in 19 patients, varied from 2.9 mV to 11.5 mV [mean 7.6 mV]. There was no hospital death. The postoperative complications included wound seroma in two patients [5%], wound hematoma in three patients [7.5%], and pacing failure in one patient [2.5%]. The subxiphoid implantation of the pacemaker using sutureless myocardial lead was valuable in obtaining safe, reliable and long-term impulse generation.