갑상선섭취율검사시(甲狀腺攝取率檢査時) 측정조건(測定條件)에 관(關)한 조사연구(調査硏究)

Study on Measurements in Thyroid Uptake Rate Test

  • 경광현 (신흥실업전문대학 방사선과) ;
  • 김화곤 (지산간호보건전문대학 방사선과)
  • 발행 : 1981.12.30

초록

This study was conducted, during the period of 20-30th, July in 1981, to survey measurement methods in thyroid uptake rate test in Seoul city. The results were summarized as follows: 1. For the great part of nuclear medcine department, a mount of radioiodine($^{131}I$) administrated to the patients was $50-100{\mu}Ci$ in thyroid uptake rate test. 2. Distribution of scintillation, counter with crystal size of $1\frac{1}{2}inch$ was 43%, 3inch(22%), 2.5inch(14%) and $2\frac{1}{2}inch$ was 7% in RAI uptake rate test. 3. When RAI uptake rate test was performed, distribution of collimator in use was flat field type collimator(78%) in general and cylindrical type collimator was 22%. 4. High voltage applied to the P-M tube was $900{\sim}1000V$(50%) and most units provided $3{\sim}15%$ of the window range for the $^{131}I$ peak $\gamma-ray$ energy. 5. Distribution on the use of neck phantom for measurements standard solution was 57% and distribution of b filter in use for room background counts and extrathyroidal tissue was 43% and 50%. 6. The distance between the counter and the source was 25cm(58%) in measuring radioactivity of standard solution, thyroid tissue and background radioactivity count. 7. The early uptake measurements(2, 4, 6 hours) are done after administration of the radioiodine dose and also 24-hour and 48-hour uptake measurements are done in routine test.