A Numerical Study on the Blood Flow through a Disc Type Heart Valve

원판형 심장판 주위의 혈액 흐름에 대한 수직 해석

  • 박영필 (연세대학교 기계공학과) ;
  • 이신재 (연세대학교 기계공학과)
  • Published : 1981.12.01

Abstract

The recirculating flows which occur in the prosthetic heart valve have been known to cause several diseases in the human body. And the recent studies show that the shear stress at the wall of the artery is also very important factor in the formation of thrombus. And many studies knave been devoted in obtaining more information about the blood flow through the prosthetic heart valve. In this Paper, the steady axisymmetric flow through the Disc-Type Heart Valve is studied by using the numerical method. The geometry of the Disc-Type Heart Valve is simplified, and the flow is assumed to be steady axisymmetric flow. The vorticity transport equation derived from the Wavier-Stoke's equation is used as the governing equation, and the explicit finite difference method is used to obtain the steady state solution. The results for several Reynolds numbers show that the recirculating flow becomes large as the Remolds number increases. Furthermore, it can be shown that the magnitudes of the vorticity and the stresses are also increased with the Reynolds number, but there is only a little change in their configurations of distribution and in their positions of maximum values.

Keywords