Selection of Well Labelled Insulin Fractions for Radioimmunoassay Use

  • Published : 1980.06.01

Abstract

Selection methods of well labelled insulin fractions based on two different criteria were compared to establish an efficient low level RIA of insulin and to elucidate the correlation between the immunoreactivity and the charcoal-adsorptivity of the radioiodine labelled insulin. The results indicated that the selection of well labelled insulin fractions by means of a charcoal-adsorption test is inappropriate. Generally, the distribution of radioactivity antibody-bindability, and charcoal-adsorptivity of the labelled insulin was not consistent with each other. Thus. the selection should be carried out for every labelling batch to get the utmost assay reliability by antibody-bindability but not by charcoal-adsorptivity. By using the well selected labelled insulin fractions based on antibody-binding, a correct assay for a reference serum was possible, and by extending the incubation time upto 96 hrs, a sharp dose response curve could be obtained even in the range of below 5 $\mu$U/ml standard insulin doses.

효과적인 저농도 인슈린 방사면역 측정법 확립 및 표지인슈린의 면역학적 환성과 탄소분말에 대한 흡착력의 상관관계를 밝히기 위하여 이들 두 다른 기준에 기초를 둔 두가지의 최적표지 인슈린선정방법을 비교한결과 탄소분말흡착에 의한 방법은 부적당함을 알 수 있었다. 일반적으로 표지인슈린의 방사능, 항체결합능, 탄소분말흡착능등은 서로 일치하지 않았다. 따라서 측정신뢰도를 높이기 위해서는 매 표지때마다 항체결합능으로 최적표지 인슈린을 선정해야 하며 탄소분말 흡착력으로 선정하면 안된다고 본다. 항체결합능에 따라 선정된 표지 인슈린을 써서 대조혈청중의 인슈린량이 정화히 측정되었으며 정온유지시간을 96시간으로 연장함으로써 5 $\mu$U/ml 하의 표준 인슈린투여 량에 대해서도 매우 예리한 응답곡선을 얻을 수 있었다.