Optimal Annealing of Natural Beryl for Thermoluminescent Dosimetry

  • Published : 1974.03.01

Abstract

The annealing of natural beryl powder used for the thermoluminescent dosimetry was investigated in order to eliminate effect of previous exposure. Through the glow curve analysis, the optimal annealing treatment was found to be one hour annealing at 145$0^{\circ}C$, but annealing for one hour at 55$0^{\circ}C$ was sufficient for reuse of natural beryl powder. There are two distinguished glow peaks at $65^{\circ}C$ and 20$0^{\circ}C$ of heating temperature during readout process. The $65^{\circ}C$ glow peak fades away rapidly, but the 20$0^{\circ}C$ glow peak remains very stable. It is, therefore, quite feasible to use the 20$0^{\circ}C$ glow peak for thermoluminescent dosimetry.

천연 녹주석의 방사선조사에 의한 효과를 없애기 위한 열처리방법을 연구하였다. Glow곡선분석에서 최적열처리시간과 온도는 145$0^{\circ}C$에서 1시간이었고, 녹주석을 다시 사용할 시에는 145$0^{\circ}C$에서 1시간동안 열처리를 다시 할 필요는 없고 55$0^{\circ}C$에서 1시간만 열처리를 하면 된다. 145$0^{\circ}C$에서 열처리된 녹주석시료를 $^{60}$Co감마선으로 100R 조사한 후 glow곡선을 그려보니 $65^{\circ}C$와 20$0^{\circ}C$에서 glow peaks를 발견하였다. $65^{\circ}C$ glow peak는 곧 없어졌으나 20$0^{\circ}C$ glow peak는 매우 안정하였다. 따라서 열형광방사선측정에는 20$0^{\circ}C$ glow peak를 이용하여 $^{60}$Co감마선을 측정할 수 있다.