Penumbra Effect on Integral Absorbed Dose in Co-60 Teletherapy

  • Published : 1973.06.01

Abstract

Due to the Co-60 source size, the penumbra in Co-60 teletheraphy poses a serious problem, even if the extended collimators are used, Here an empirical formula for the calculation of integral absorbed dose in the penumbra region was derived. Through a numerical calculation, the penumbra effect on integral absorbed dose was investigated. The longer the source-to-skin distance, the larger the integral absorbed dose of penumbra region, and the larger the source diameter, the larger the integral absorbed dose of penumbra region. It was also found that in some case the integral absorbed dose in penumbra region becomes several times larger than the integral absorbed dose of treatment region itself if the source-to-skin distance becomes greater. Therefore, one must consider the penumbra effect in Co-60 teletherapy.

Co-60원격치료장치에 장진된 Co-60 선원의 크기 때문에 Co-60원격치료시 치료부위에 반영부분이 생기게 되어 불필요한 방사선조사를 받게된다. 반영부분에서의 흡수적 산선량을 계산할 수 있는 식을 실험적으로 유도하였고, 실재로 Co-60원격치료에서 사용되는 조건을 실험식에 대입하여 반영의 크기가 흡수적산선랑에 미치는 영향을 연구해 보았다. 선원과 표피간의 거리가 크면 클수록 반영부분의 흡수적산선량은 커지고, 선원의 직경이 크면 클수록 반영부분의 흡수적산선량이 커진다. 선원과 피부간의 거리가 커지는데 따라, 어떤 경우에는 반영부분의 흡수적산선량이 치료부분의 흡수적산선량의 수배가 될 때도 있음을 알았다. 따라서 Co-60원격치료시에는 반영효과를 고려에 넣어 불필요한 방사선조사를 피하도록 하여야 하겠다.