An Experimental Approach for Verifying the Effect of Scattered Gamma-rays on the “Before Glow”in a Thermoluminescent Glow Curve

  • Published : 1972.03.01

Abstract

In order to verify the contribution of scattered photons in a restricted gamma-cell as a cause of the“before glow”on a thermoluminescent glow curve of natural quartz, the ratio of the scattered to primary radiation contributions (S/P) in the cell is measured and the relationship between the effective“before glow”height ( $h_{b}$) and S/P ratio is quantitatively investigated. The result shows quite good linear relationship between them with a correlation coefficient of +0.9, which possibly suggests that the electrons originally released by the photons of reduced energy are trapped in the shallower traps. Moreover, the ratios of $h_{b}$ to total glow area (At) and of effective “before glow”area (Ab) to At are also examined to see the relationships between S/P and each of them, respectively. The relationships are represented by exponential functions in the region of S/P greater than 0.035. Finally, the exposure limit for re-use of the natural quartz as a TLD was found to be approximately 10$^{5}$ R by analyzing total thermoluminescent output and corresponding exposure dose.ose.

수정의 영형광 glow curve 상,"before glow"의 생성원인중의 하나가 밀폐된 제한 공간내의 산란 감마선의 기여임을 확인하기 위하여 산란선대 일차선의 기여비(S/P)를 측정하였다. 이 S/P와 "before glow"의 유효높이 ($h_{b}$)와의 상관 관계를 고찰하였는바 상관계수+0.9라는 비교적 밀접한 일차 관계가 있음을 알았으며 이는 에너지가 감소된 산란 선에 의하여 여기되었던 전자가 본래 일차선으로 여기되었던 전자보다 얕은 trap에 걸려 있었음을 입증하는 것으로 보인다. 한편 $h_{b}$와 glow curve의 전면적 (At)의 비 및 "before glow"의 유효면적(Ab)과 At 와의 비가 S/P와 어떤 관계에 있는가도 조사하였는데 이들의 관계는 단순하지 않으며 다만 S/P 값이 0.035 보다 큰 영역에서는 간단한 대수함수로 표현되었다. 끝으로 자연수정을 TLD로 사용할 경우 그 재사용을 위한 선량한계를 살펴본 결과 그것은 $10^{5}$ R 정도임을 알아내었다.