Abstract

A neutron detector, in general, can not be utilized as the thermal neutron detecting chamber in the nuclear power reactor, especially P.W.R. due to the characteristics of high temperature, high pressure and high neutron flux in a reactor vessel. We have performed an experiment to detect the thermal neutrons at 400.deg. C and high flux of thermal neutron in a power reactor. Coating boron-10 on the aluminium plates by means of surface diffusion method at 600.deg. C for 5 hours in an electric furace, also we made a typical chamber which was compensated ionization chamber filled with free air as an ionization gas. It was checked the chamber characteristics in the TRIGA MARK-II Reactor at the power level from zero to 250KW. The chamber current showed a perfect linear increase to power increase. However, many variation of the measured current were observed within the power of 50KW.

Keywords