Proceedings of the KIEE Conference (대한전기학회:학술대회논문집)
- 2015.07a
- /
- Pages.1431-1432
- /
- 2015
Assessment of efficacy of respiration monitoring system customized for patients under critical care
위급 중환자에 특화된 호흡 모니터링 시스템의 유효성 평가
-
Lee, In-kwang
(Biomedical Engineering Department, School of Medicine, Chungbuk National University) ;
-
Park, Mi-Jung
(Biomedical Engineering Department, School of Medicine, Chungbuk National University) ;
- Kim, Kyoung-Oak (Department of Nursing, Woosong College) ;
-
Cha, Eun-Jong
(Biomedical Engineering Department, School of Medicine, Chungbuk National University) ;
-
Kim, Kyung-Ah
(Biomedical Engineering Department, School of Medicine, Chungbuk National University)
-
이인광
(충북대학교 의과대학 의공학교실) ;
-
박미정
(충북대학교 의과대학 의공학교실) ;
- 김경옥 (우송정보대학 간호과) ;
-
차은종
(충북대학교 의과대학 의공학교실) ;
-
김경아
(충북대학교 의과대학 의공학교실)
- Published : 2015.07.15
Abstract
A system was developed to continuously monitor the respiratory signal under critical care, followed by performance evaluation for clinical application. Very much accurate pressure measurement was made in the endo-tracheal tube with mean relative error of 0.13%. In the comparative experiment simulating inspiration and expiration, the tidal volume measurement was also accurate within mean relative error of 3%, validating reliable measurement of respiratory signals.
Keywords