Near Infrared Wavelength Calibration without Telluric OH Lines

  • Nguyen, Ngan N.K. (School of Space Research, Kyung Hee University) ;
  • Pak, Soojong (School of Space Research, Kyung Hee University) ;
  • Kang, Wonseok (School of Space Research, Kyung Hee University)
  • 발행 : 2012.10.17

초록

Grating spectrograph observation in near-infrared bands requires wavelength calibrations. We first need to extract order-strips from the echellogram data using flat images, and then to correct the spatial distortions and wavelength scales of the order strips using calibration arc lamps or the telluric OH emission lines. For very high resolution spectrograph using echelle, however, the arc lamps do not have enough emission lines in the order-strips, so we usually use the OH lines for accurate wavelength calibrations in wavelengths shorter than 2.3 microns. In this poster, we present wavelength calibration methods for longer wavelength bands using the telluric absorption lines in early-type stellar spectra and the telluric atmospheric transmission models. This technique will be applied to the data reduction pipeline for the IGRINS.

키워드