Radiation Quality Analysis of Diagnostic X-ray Equipment

진단용 X선 기기의 선질 분석

  • Published : 2010.10.27

Abstract

X-ray equipment used to diagnose a patient has a big defect of a patient's exposure to radiation caused in irradiating X-ray to the human body, ICRP restricts the permissible exposure dose of the human body. A number of studies has been proceeded to reduce these exposures. In this study the high voltage generator with inverter system, which is possible to increase the generation efficiency of X-ray and to control the precise output power was produced. Also, to minimize the ripple which is possible to occur in the direct voltage that is applied to X-ray tube the propagation rectification method was applied and the radiation reproducibility and properties were evaluated to use this for the diagnosis of patient.

환자의 진단에 유용하게 사용되고 있는 X선 장치는 X선을 인체에 조사할 때 나타나는 피폭이 큰 단점으로 지적되고 있으며 ICRP에서는 인체의 방사선 노출 허용량을 제한하고 있다. 이러한 피폭을 줄이기 위해 많은 연구 개발이 진행되고 있다. 본 연구에서는 X선의 발생 효율 증가 및 정밀한 출력 제어가 가능한 인버터 방식의 고전압 발생장치를 제작하였다. 또한 X선관에 인가되는 직류 고전압에 발생 가능한 리플을 최소화하기 위해 전파 정류방식을 적용하였으며, 이를 환자의 진단에 사용하기 위해 조사재현성 및 선질을 평가하였다.