Appropriate image quality management method of bone mineral density measurement

골밀도 측정의 올바른 질 관리방법

  • Kim, Ho-Sung (Department of Nuclear Medicine, Asan Medical Center) ;
  • Dong, Kyung-Rae (Department of Radiological Technology, Gwangju Health College University)
  • Published : 2009.05.22

Abstract

In Bone Mineral Density(BMD) measurements, accuracy and precision must be superior in order to know the small changes in bone mineral density and actual biological changes. Therefore the purpose of this study is to increase the reliability of bone mineral density inspection through appropriate management of image quality from machines and inspectors. For the machine management method, the recommended phantom from each bone mineral density machine manufacturer was used to take 10~25 measurements to determine the standard amount and permitted limit. On each inspection day, measurements were taken everyday or at least three times per week to verify the whether or not change existed in the amount of actual bone mineral density. Also evaluations following Shewhart control chart and CUSUM control chart rules were made for the bone mineral density figures from the phantoms used for measurements. Various forms of management became necessary for machine installation and movement. For the management methods of inspectors, evaluation of the measurement precision was conducted by testing the reproducibility of the exact same figures without any real biological changes occurring during reinspection. There were two measurement methods followed: patients were either measured twice with 30 measurements or three times with 15 measurements. An important point to make regarding measurements is that after the first inspection and any other inspection following, the patient was required to come off the inspection table completely and then get back on for any further measurements. With a 95% confidence level, the precision error produced from the measurement bone mineral figures produced a precision error of 2.77 times the minimum of the biological bone mineral density change (Least significant change: LSC). In order to assure reliability in inspection, there needs to be good oversight of machine management and measurer for machine operation and inspection error. Accuracy error in machines needs to be reduced to under 1% for scientific development in bone mineral density machines.

골밀도 측정은 정확도와 정밀도가 우수하여야 작은 골량의 변화에도 진정한 생물학적 변화를 알 수 있다. 따라서 장비 및 검사자의 올바른 질 관리를 통하여 골밀도 검사의 신뢰성을 높이는 것을 목적으로 한다. 장비관리방법은 각각의 골밀도 장비 제조사에서 권고하는 팬텀을 이용하여 10~25회 측정하여 기준 값과 허용 범위를 정하고 검사가 있는 날에 매일 측정하거나 일주일에 3회 이상 측정하여 실제 골밀도의 값의 변화 유무를 확인하여야 한다. 또한 측정된 팬텀의 골밀도 수치를 기록하여 Shewart control chart와 CUSUM control chart를 만들어 각각의 Rule에 따라 평가한다. 이러한 관리는 장비의 설치 및 이동 시에 반드시 행해져야 한다. 검사자 관리방법은 정밀도 측정으로 평가하는데 정밀도는 재검사하였을 때에 실제 생물학적의 변화 없이 수치상의 결과값을 똑같이 재현할 수 있는지 알아보는 것이다. 측정 방법은 환자를 두 번씩 30번 측정하는 방법과 세 번씩 15번 측정하는 방법이 있다. 측정에서 중요한 것은 한 번 검사 후 두번째나 세 번째 검사에서도 반드시 검사 테이블에서 내려왔다 다시 올라가서 검사를 해야 한다. 측정된 골밀도 수치를 이용하여 정밀오차를 산출하여 95% 신뢰수준으로 정밀오차에 2.77을 곱하여 최소한의 생물학적 골밀도 변화를 산출한다. 골밀도 장비는 과학의 발달로 인하여 장비의 정확 오차가 1%이내로 줄었기 때문에 장비관리와 측정자의 기계조작 및 검사 오차를 잘 관리한다면 검사의 신뢰성 확보에 도움을 줄 것이다.

Keywords