Setup and Uncertainty of Standard System for Calibrating Pulse Generator of Partial Discharge

부분방전 교정펄스 발생기의 표준화 시스템구축과 불확도

  • Kim, K.H. (Korea Electrotechnology Research Insititute) ;
  • Yi, S.H. (Korea Electrotechnology Research Insititute) ;
  • Lee, H.J. (Korea Electrotechnology Research Insititute) ;
  • Kang, D.S. (Korea Electrotechnology Research Insititute)
  • Published : 2009.07.14

Abstract

This paper describes the standard system for calibrating pulse generator of partial discharge(PD) and its uncertainty. The system is consisted of digital pulse generator, digital recorder and evaluation software. The uncertainty requirement of calibrator charge is less than $\pm$ (0.1pC + 0.03q) and that of pulse duration is less than $\pm$ 10 %. The system can generate various kind of calibration pulses such as single pulse, double pulses, oscillation pulse, long-duration pulse, random pulses and evaluate their uncertainty.