Abstract

In this paper, we designed bio-signal acquisition system in Magnetic Resonance Imager(MRI) Environment. In MRI Environment, Strong RF Pulse and Gradient Field Switching Noise exist and can cause distortion of ECG. By this, ECG can lose their important information. So we proposed a bio-signal acquisition system with robust immunity to RF pulse and gradient switching noise. In conclusions, the proposed system showed the prevent saturation of measured biosignal and possibility of using cardiac gating and respiration gating method.

Keywords