The Effects of AFM Microcantilever Characteristics on the Non-Contact Mode Measurements
AFM 마이크로캔틸레버 특성에 따른 비접촉모드의 영향 고찰
Abstract
In non-contact mode atomic force microscopy, the response of a resonating tip is used to measure the nanoscale topography and other properties of a sample surface. However, the tip-surface interactions can affect the tip response and destabilize the non-contact mode control. Especially it is difficult to obtain a good scanned image of high adhesion surfaces such as polymers using conventional hard NCHR tip and non-contact mode control. In this study, experimental investigation is made on the non-contact mode imaging and we report the microcantilever having low stiffness (OMCL) is useful to measure the properties of samples such as elasticity. In addition, we proved that it was adequate to use low stiffness microcantilever to obtain a good scanned image in AFM for the soft and high adhesion sample.