Design of Readout IC for Uncooled Infrared Bolometer Sensor using Bias Offset Correction Technique

오프셋 보정 기술을 이용한 비냉각형 적외선 센서용 신호검출 회로 설계

  • Published : 2005.07.07

Abstract

Infrared bolometer sensor's variation is detected by voltage drop between reference resistor and bolometer resistor in this architecture. One of the serious problems in this architecture is that these resistors value has a process variation. So common-mode level could be different from expectation in room temperature. Different common-mode level could lead to wrong output at the end of readout circuit. We suggest useful method to solve this problem. Difference correction using capacitor has reduced CM level difference to 86% for 1 $M\Omega$. bolometer and reference resistor's 10% variation.

Keywords