Abstract

In the cardiovascular system, the waveform of the pulsatory blood pressure appears variously due to the cardiac impulse and compliance of blood vessels and arm tissue. We have constructed a blood pressure simulator to investigate effects of mechanical properties of artery walls and tissue on blood pressure measurements. The blood pressure simulator is designed to reproduce wave forms of blood pressure in human arteries. To minimize tracking error, we use a linear control valve, and adapt a hybrid control scheme which consists of a feedback controller and a feedforward controller. Any form of the pressure wave can be reproduced, changing function of the wave form in the computer connected to the simulator for control. From experiments, it has been shown that the simulator reproduces wave forms very well, and that the hybrid scheme adapted is superior to the feedback controller.