Determination of Proton Beam Position Based on Prompt Gamma Ray Detection

즉발감마선을 이용한 양성자 빔 위치 측정에 관한 연구

  • Published : 2004.11.26

Abstract

The proton therapy of radiation therapy methods using Bragg Peak which is proton beam's characteristic dose distribution can give a normal tissue lower dose than cancer, comparing with the former existing radiation therapy methods. For exact treatment and patient' safety, we need to know proton beam's position in body, but a proton beam completely stops at treatment region and proton beam's range is uncertainly made by the variety of organs having each different density, so we aren't able to find a proton beam' position by suitable methods yet. With Monte Carlo Computing Method, as a result that we had simulated prompt gamma detection system using correlation of proton beam's absorbed dose distribution about water and prompt gamma distribution by nuclear interaction occurred by collisions of proton and water's hydrogen atoms, we could confirm that a proton beam's position was able to detect by using simulated prompt gamma detection system in body on the real-time

양성자 치료는 양성자 빔의 선량분포 특성인 브래그 피크를 이용한 방사선 치료방법으로 기존의 전자 및 광자빔을 이용하는 방사선 치료와 비교하여 정상세포에는 현저히 적은 선량을 주고, 암세포에는 대부분의 선량을 줄 수 있어 최근에 각광받고 있다. 양성자 치료 빔은 치료 부위에서 완전히 멈추고 인체 내 각기 다른 밀도를 가진 장기들에 의해 양성자 빔의 비정이 불확실하기 때문에 체내 양성자 빔의 분포를 확인하고 빔을 조정하면서 치료할 수 있는 적절한 방법이 없다. 양성자 빔의 물질에 대한 흡수선량 분포와 물질과 양성자의 핵반응으로 방출되는 즉발 감마선의 상관관계를 이용한 측정시스템을 몬테칼로 전산코드로 전산모사하고 계산한 결과 체내 양성자 빔의 위치를 실시간으로 측정할 수 있음을 확인하였다.