Comparison of Signal to Noise Ratios in different configulation of optical coherence tomography systems

간섭계 형태에 따른 광단층 촬영장치의 SNR 비교

  • Published : 2004.05.01

Abstract

As a noninvasive imaging meathod, optical coherence tomography system has been extensively studied because it has some advantage such as imaging of high resolution, low cost, and compace size configuration. The optical power of the sample signal reflected from an object is usually very small in optical coherence tomography(OCT) due to absorption and scattering in a sample material. Furthermore, there happens a serious loss of sample signal power in a fiber coupler because it does not transfer to a photodetector but return to the optical source. In order to improve the SNR of OCT system, basic Michelson and Mach-Zehnder interferometer types were configured then, we compared simulation with measurement of reference sample Therefore, an On in configuration of Math-Zehnder interferometer was demonstrated in order to improve the signal to noise ratio

인체에 무해하며 고분해능의 단층영상과 저가격 및 소형으로 제작이 가능하여 전세계적으로 많은 연구가 진행중인 광영상 단층촬영기(Optical coherence tomography system : OCT)에 관한 것이다. 샘플에서 흡수와 산란으로 인하여 샘플로부터 반사된 광신호는 매우 미약하다. 게다가 커플러등에 의한 손실로 인하여 광검출기에 도달하는 신호는 매우 미약하다. 이러한 On시스템의 신호대 잡음비를 향상시키기 위하여 기본적인 마이켈슨 간섭계와 마흐젠더 간섭계를 구성하여 시뮬레이션과 기준샘플을 이용하여 측정하였다. 실험결과 마흐젠더 간섭계를 사용한 샘플로 반사된 광신호는 OCT 시스템의 신호대 잡음비 향상시키기 위하여 두 가지 형태의 간섭계를 구성하였으며 시뮬레이션을 통한 비교 분석과 기준 샘플의 측정에 의한 비교분석을 하였다. 실험결과 마흐젠더 간섭계형 OCT 시스템이 신호대 잡음비가 향상됨을 확인 하였다.