MTF 측정을 통한 비정질 셀레늄 기반의 디지털 방사선 검출기의 영상 질 평가에 관한 연구

Image Quality of Amorphous Selenium DR system using MTF measurement

  • Seok, Dae-Woo (Department of Biomedical Engineering of Inje University) ;
  • Park, Ji-Koon (Department of Biomedical Engineering of Inje University) ;
  • Choi, Jang-Yong (Department of Biomedical Engineering of Inje University) ;
  • Nam, Sang-Hee (Medical Imaging Research center of Inje University) ;
  • Kang, Shin-Won (School of Electrical Engineering and Computer Science of Kyungpook National University)
  • 발행 : 2003.11.13

초록

In this paper, the evaluation of image quality was performed for digital radiography which is developing in using amorphous selenium as a photoconductor material for the purpose of offering basic research data and measurement technique about Medical Imaging Quality. So Modulation Transfer Function as a main factor of imaging quality evaluation was investigated by slit method. For measurement of MTF, Nuclear associates. 07-624 Slit camera image was obtained to study the variation of MTF corresponding to changing spatial frequency. And Presampling MTF was estimated by slit camera image with $10\;{\mu}m$ width at Digital Radiography. In this study, the obtained data demonstrates that the clinical value of a direct conversion type digital radiation detector using the amorphous selenium, which is being developed by domestic technology.

키워드