Abstract
A glassy carbon electrode(GCE) modified with nafion-DTPA-glycerol was used for the highly selective and sensitive determination of a trace amount of Cu$\^$2+/. Various experimental parameters, which influenced the response of nafion-DTPA-glycerol modified electrode to Cu$\^$2+/, were optimized. The copper(II) was accumulated on the electrode surface by the formation of the complex in an open circuit, and the resulting surface was characterized by medium exchange, electrochemical reduction, and differential pulse voltammetry, A linear range was obtained in the concentration range 1.0${\times}$10$\^$-8/M∼1.0${\times}$10$\^$-6/M Cu(II) with 7 min preconcentration. Further, when an approximate amount of lead(II) is added to the test solution, nafion-DTPA-glycerol modified glassy carbon electrode has a dynamic range of 2 orders magnitude(1.0${\times}$10$\^$-9/M∼1.0${\times}$10$\^$-7/M). The detection limit(3 $\sigma$) was as low as 5.0${\times}$10$\^$-6/M(0.032ppb). The interferences from other metal ions could be reduced by adding KCN into the sample solutions. This method was applied to the determination of coppe,(II) in certified reference material(3.23${\times}$10$\^$-7/M, 21ppb), sea water(9.50${\times}$10/sup-7/M, 60ppb). The result agrees satisfactorily with the value measured by Korea Research Institute of Standard and Science.