Evalution of Hemolysis in Axial Flow Blood Pump with Computational Fluid Dynamics Analysis

전산유체해석을 이용한 축류형 혈액펌프의 용혈평가

  • 임상필 (순천향대학교 대학원 정보제어공학과) ;
  • 김동욱 (순천향대학교 대학원 정보제어공학과)
  • Published : 2003.06.01

Abstract

Artificial heart is divided pulsation style and nonpulsation style greatly according to flowing of blood. nonpulsation pump is advantage of miniaturization avaliable because it is simple and non-volumic-pump than pulsation pump. Non pulsation pump is derided axial flow style and centrifugal style accordig to rotating style. An axial flow blood pump can be made smaller than a centrifugal blood pump because of its higher specific speed. A hemolysis is an important factor for the development of an axial flow blood pump. It is difficult to identify the areas where hemolysis nun. Evaluation of hemolysis both in in vitro and in vivo require a long time and are costly. Computational fluid dynamics(CFD) analysis enables the engineer to predict hemolysis on a computer. The aims of this study is Computational fluid dynamics in the whole axial flow pump and to verify the accuracy of prediction results of CFD analysis compare with in vitro experimental results.