Energy Dependancy of the Polyethylene Terephthalate Film for Radiation Detector

방사선 검출기용 PET 박막의 에너지 의존성

  • Published : 2002.11.07

Abstract

Currently small and accurate dosimeters are on the rise. In this study, the feasibility and energy dependency of the electret dosimeter that made of PET (polyethylene terephtalate) were observed by irradiating 4, 6, 15 MV photon beams from the clinical linear accelerator to develop a dosimeter for the clinical field. $10{\times}10cm$ field size of the photon beams were irradiated to the electret dosimeter where the 2.5 cm depth in the polystylene phantom from 100 cm SSD, while 300 DCV was applied to the electret dosimeter. The result showed that the absorbed dose was proportional to the charge linearly, and the volume of a dosimeter could be reduced and the signals were high enough. According to this study, it was found that the polymer electret detector could be produced as a large quantity with a small cost and showed the feasibility of a realtime measurement.