Cross Capacitance 원리를 작용한 새로운 유전상수 측정방법 제안

A New Measurement Method of Dielectric Constants Applied the Principles of Cross Capacitance

  • 김한준 (한국표준과학연구원 전기.자기그룹) ;
  • 이래덕 (한국표준과학연구원 전기.자기그룹) ;
  • 강전흥 (한국표준과학연구원 전기.자기그룹) ;
  • 유광민 (한국표준과학연구원 전기.자기그룹) ;
  • 한상옥 (충남대학교 전기공학과)
  • 발행 : 2002.07.08

초록

The guard-ring type 3-terminal parallel plate electrodes proposed by ASTM D 150-81 and IEC 250 have been widely used for measurement of dielectric constants of solid dielectrics. However the method using this electrodes causes many uncertainty associated with the measurement errors of the diameter of the guarded electrode. the gap between guarded and guard-ring electrode. the distance of two active electrodes(the thickness of specimen), the roughness and contamination of surface of electrode and specimen. close adherence grade of electrode and specimen. In this paper. a new electrode system of cross capacitance type based on Thompson-Lampard theorem is designed and is employed for the measurement of dielectric constant. The results of simulation of guard-ring electrode and cross capacitance electrode using FEM program show that distance measurement between two electrodes in guard-ring electrode produces large uncertainty. on the other hand this effect in cross capacitance electrode is negligible. Furthermore. the air gap effects in the cross capacitance electrode is 5.6 times less sensitive than that in guard-ring electrode by assuming air gap of $50{\mu}m$.

키워드