Development of a high-performance liquid chromatographic method for the determintion of levosulpiride in plasma

  • Published : 2002.10.01

Abstract

Levosulpiride is the levo-enantiomer from of racemic sulpride. abenzamide derivative selectively inhibition doparninergic D2 receptos at the trigger zone both in the central nervous system and in the gastrointestinal tract. We report a rapid and sensitive HPLC method using reverse phase C 18 column with fluorescence detection for separation and quantitation of levosulpiride in plasma. Tiapride was used as an internal standard. After adding an internal standard. levosulpiride in 800 ${\mu}l$ of plasma was extracted under basic conditions with ethyl acetate and methylene chloride. (omitted)