Precise Measurement of Beam Energy and Range with TOF and Counter Telescope System

  • Nanbu, Shuya (Department of Energy Sciences, Tokyo Institute of Technology, Yokohama) ;
  • Kanai, Tatsuaki (Research Center of Charged Particle Therapy, National Institute of Radiological Sciences, Chiba) ;
  • Kohno, Toshiyuki (Department of Energy Sciences, Tokyo Institute of Technology, Yokohama) ;
  • Ohno, Yumiko (Department of Energy Sciences, Tokyo Institute of Technology, Yokohama)
  • Published : 2002.09.01

Abstract

In order to improve the accuracy of charged-particle radiation therapy, the beam energy was measured precisely using a TOF-system, and the range using a counter telescope system. A Si detector and a Ge detector were used to estimate the range straggling as a $\Delta$E and an E detector, respectively, because they have good energy resolution and the output pulse heights don't depend on the atomic number of detected particles. The results were compared with the theoretical values by a calculation code.

Keywords