Verification of Periodical Calibration for Iso-center Positions using Quality Assurance System for Irradiation Equipment Position Established at PMRC

  • Yasuoka, Kiyoshi (Proton Medical Research Center(PMRC), University of Tsukuba, Japan) ;
  • Ishikawa, Satoko (Proton Medical Research Center(PMRC), University of Tsukuba, Japan)
  • 발행 : 2002.09.01

초록

We present the results on the calibration of iso-center positions using the quality assurance system established at PMRC for determination of center position in X-ray and proton irradiation fields. Details on the system are presented in another presentation in this session. The equipment in the system is mounted on a patient treatment bed in each proton exposure room, G1 or G2. A center of a stainless ball on the equipment is set at a cross of laser markers located around the iso-center and fixed on the room and on the snout in the gantry. A proton beam or an X-ray beam is exposed onto the ball through a brass collimator of 100 mm ${\times}$ 100 mm and projected onto the imaging plate set at I cm behind the ball. On the axis perpendicular to the thrust axis of the gantry on the imaging plate, a distance between a center of the collimator image and a center of the ball image varies as a cosine function of gantry angles unless the ball is set on the iso-center. An amplitude of the cosine curve shows the distance between the ball and the iso-center, an offset the offset of the collimator, and a phase shift at a zero crossing point the ball direction viewed from the iso-center. We present the relation among the iso-center position, the laser maker position, and the center of proton and X-ray irradiation fields. Its stability and its reproducibility are discussed.