The Study on Pre-treatment Method of Filter for Analysing the Heavy Metals in Air Quality

대기중의 중금속 분석을 위한 여지의 전처리 방법에 관한 연구

  • 김광래 (서울시보건환경연구원 연구기획팀) ;
  • 이상칠 (서울시보건환경연구원 연구기획팀) ;
  • 어수미 (서울시보건환경연구원 연구기획팀) ;
  • 김민영 (서울시보건환경연구원 연구기획팀) ;
  • 신재영 (서울시보건환경연구원 연구기획팀) ;
  • 이재영 (서울시립대학교 환경공학부)
  • Published : 2002.04.01

Abstract

This study was carried out to establish the standardized analysis method in order to decide accurate concentrations of hazardous metals in the air. Acid decomposition method used usually was compared to microwave digestion method. Comparing results of tested background concentration in blank filter, we found that the magnitude of element concentration was Na, Ca, K, Mg, Al and Ba by order. The element concentrations of Glass fiber filter were higher than those of Quartz fiber filter, and the number of undetected components in Glass fiber filter was lower than that in Quartz fiber filter. Thus it is supposed that the concentration of background elements in Glass fiber filter was higher than those in Quartz fiber filter The extraction rate of microwave digestion method was superior to those of acid decomposition method during the test of blank filter as well as SRM of NIST. In case of the SRM of NIST, the average extraction efficiency of acid decomposition Method and microwave pretreatment is 53.8∼82.7%, 81.3∼97.1%, respectively. This result might be caused by the closed system of Microwave, which make outflow and loss of components less. Also microwave digestion method has other merits such as the minimization of time, reagents, and contamination. Furthermore, if the extraction condition, extraction time and used acids are optimized, the better results will be represented.

Keywords