유체기계공업학회:학술대회논문집
- 유체기계공업학회 2002년도 유체기계 연구개발 발표회 논문집
- /
- Pages.378-386
- /
- 2002
간접 교정에 의한 다회선 초음파유량계 UR-1000 불확도 분석
Uncertainty Analysis for the Multi-path Ultrasonic Flowmeter UR- 1000 with Dry Calibration
- 발행 : 2002.12.01
초록
Multi-path ultrasonic Sow measurement system uncertainty is determined by assigning an expected error of each component of flow measurement and then defining the total flow measurement uncertainty as square root of the sum of squared values of the individual error. Sources of uncertainty for flow measurement are geometry, transit time and velocity profile integration uncertainty. A theoretical uncertainty model for multi-path ultrasonic transit time flowmeter configured with parallel 5 chords, is derived from and calculated by dry calibration method.
키워드