의료용 펌프의 리니어 모터 설계 및 해석

Design and Analysis of Linear Motor for Medical Appliance Pump

  • 발행 : 2002.07.10

초록

In this paper, it has been investigated that the design and simulation of linear motor for medical appliances. Vacuum pump consists of unified plunger and piston part, coils, inlet valve and outlet valve. Operating principal of vacuum pump show that air is to flow to go inside through the inlet valve, when the magnetic force (16N) is exerted, whereas, air is to flow to go outside through outlet valve, when spring force (14.8N) is exerted. The results of simulation, it was turned out that magnetic force was loaded with 16 Newton, displacement of actuator was about 8mm.

키워드