대한전기학회:학술대회논문집 (Proceedings of the KIEE Conference)
- 대한전기학회 2001년도 춘계학술대회 논문집 전기기기 및 에너지변환시스템부문
- /
- Pages.22-25
- /
- 2001
중자장급 보급형 국산 초전도 MRI 마그네트 개발
The Development of Popular type Domestic Superconducting MRI Magnet with Middle Magnetic Field Range
초록
The research results on the superconducting magnet for whole body MRI are presented. The magnet consists of main coil with 6 solenoid coils, shielding coil with 2 solenoid coils and 6 sets of cryogenic shim coil. The ferromagnetic shim assembly is installed on the inside wall of the room temperature bore for shimming inhomogeneous field components generated due to manufacturing tolerances, installation misalignments and external ferromagnetic materials near the magnet. Also, the magnet is enclosed with the horizontal type cryostat with 80cm room temperature bore to keep the magnet under the operating temperature. The magnetic field distributions within the imaging volume were measured by the NMR field mapping system. Through the test, the central field of magnet was 1.5 Tesla and the field homogeneity of 9.3 ppm has been obtained on 40cm DSV(the diameter of spherical volume) and using this magnet, comparatively good images for human body, fruits and water phantoms have been achieved.
키워드