인공고관절 생체재료 마멸평가를 위한 시뮬레이터 개발

Development of Hip Joint Simulator to Evaluate The Wear of Biomaterials Used in Total Hip Joint Replacement

  • 이권용 (대구대학교 자동차·산업·기계공학부) ;
  • 윤재웅 (대구대학교 자동차·산업·기계공학부) ;
  • 전승범 ((주)알앤비) ;
  • 박성길 ((주)알앤비)
  • 발행 : 2001.06.01

초록

Hip joint simulator which Is an essential device for evaluating the wear of biomaterials (ultrahigh molecular weight polyethylene, Co-Cr alloy, alumina, etc.) used in total hip joint replacement was developed. This hip joint simulator mimics the joint motion and joint loading of human gait by adapting the 4 degree of freedom in kinematic motion (flexing/extension, adduction/abduction, Internal rotation/external rotation) and axial loading, Four stations are operated by 8 servo-motors and harmony drives. Joint leading was imposed by displacement control from a ball screw, LM guide, and spring system. Each kinematic link system operates separately or coupled modes. A heater and a thermocouple were installed for keeping the body temperature in each station.