Calibration of the Broadband Sensor(STS-2) using the Step Method

스텝방법을 이용한 광대역지진계 센서(STS-2)의 검증

  • 류용규 (기상청 지진담당관실) ;
  • 이덕기 (기상연구소 해양지진상연구실) ;
  • 이전희 (기상청 지진담당관실) ;
  • 오석훈 (기상연구소 해양지진상연구실)
  • Published : 2001.09.01

Abstract

A close examination of the step calibrations for STS-2 seismometers installed by KMA has been done to deduce the response parameters of those instrument including angular corner frequency, damping factor and coil constant factor. A non-linear least square inversion method has been apple iud to estimate these parameters. The estimated parameters coincide with the manufactory specification with less than 1% error. This method will be extended near- future to deduce the response parameters for SS-1 short period seismometer.

Keywords