Magnetic Impact Actuator for Robotic Endoscope

대장내시경을 위한 자기 충격 액츄에이터

  • 민현진 (한국과학기술원 기계공학과) ;
  • 임형준 (한국과학기술원 기계공학과) ;
  • 김병규 (한국과학기술원) ;
  • 김수현 (한국과학기술원 기계공학과)
  • Published : 2001.04.01

Abstract

For robotic endoscope, some researchers suggest pneumatic actuators based on inchworm motion. But, the existing endoscopes are not seemed to be replaced completely because human intestine is very sensitive and susceptible to damage. We design and test a new locomotion of robotic endoscope able to maneuver safely in the human intestine. The actuating mechanism is composed of two solenoids at each side and a single permanent magnet. When the current direction is reversed, repulsive force and attractive at the opposition side propels permanent magnet. Impact force against robotic endoscope transfer momentum from moving magnet to endoscope capsule. The direction and moving speed of the actuator can be controlled by adjusting impact force. Modeling and simulation experiments are carried out to predict the performance of the actuator. Simulation experiments show that force profile of permanent magnet is the dominant factor for the characteristic of the actuator. The results of simulations are verified by comparing with the experimental results.