A Study on an Implementation of Bioelectric Impedance Instrumentation

생체 임피던스 계측 방법의 구현에 관한 연구

  • 김홍석 (한국생산기술연구원 첨단생산설비연구팀) ;
  • 박세화 (한국생산기술연구원 의료반도체기술연구팀)
  • Published : 2000.10.01

Abstract

A new bioelectric impedance measurement method is proposed for the precise measurement of the bioelectric impedance. To obtain the impedance from the known applied a.c. current and the measured voltage signals, a frequency conversion circuit, like the mixer in heterodyne receivers, is introduced to reduce the frequency of the original current. It can be observed from several lines of derivation that the impedance is independent on the amplitude and phase of the mixing signals. This makes it possible to use low-speed analog-to-digital converters and thus utilize cheaper electronic parts in the implementation. The possibility of the method is shown by simulations, and a generic structure applicable to bioelectric impedance measurement devices is also proposed.

Keywords