Effect of Coincidence Gamma-ray Spectroscopy to the Reduction of Background Spectrum

  • Published : 1998.05.01

Abstract

A coincidence gamma-ray spectroscopy method was applied to reduce the background radioactivity for measuring the activity of radioisotopes in a sample in the presence of environmental natural radioactivity. A HPGe detector was used for the coincident spectrum as a main detector and a NaI(Tl) scintillation detector for gating purposes as an associated detector. For coincidence spectroscopy the whole energy spectrum of associated detector was used instead of gate signals. The coincident events obtained from the gating spectrum was evaluated by a coincidence computer program in this study instead of timing circuit. In this work, the background of detection environment was reduced to factor 100 and peaks to be determined was reduced to factor 30 using the coincidence gamma-ray spectroscopy.

Keywords