Proceedings of the KOSOMBE Conference (대한의용생체공학회:학술대회논문집)
- Volume 1997 Issue 05
- /
- Pages.108-110
- /
- 1997
Minimum-Power and/or Minimum-Inductance Design of MRI Gradient Coils Using Loop-Current Elements
환전류소를 이용한 MRI용 경사자계코일의 최소전력/최소인덕턴스설계
- Lee, D.R. (Department of Electronics and Info. Engr., Korea University) ;
- Yang, Y.J. (Department of Electronics and Info. Engr., Korea University) ;
- Kim, S.K. (Department of Electronics and Info. Engr., Korea University) ;
-
Ahn, C.B.
(Kwang-Woon University) ;
-
Lee, H.K.
(Medison Ltd.) ;
-
Oh, C.H.
(Department of Electronics and Info. Engr., Korea University)
- 이덕래 (고려대학교 전자및정보공학부) ;
- 양윤정 (고려대학교 전자및정보공학부) ;
- 김선경 (고려대학교 전자및정보공학부) ;
-
안창범
(광운대학교) ;
-
이흥규
((주)메디슨) ;
-
오창현
(고려대학교 전자및정보공학부)
- Published : 1997.05.23
Abstract
In MRI, gradient coils are needed for spatial selection and position coding to obtain the position information of the NMR signal. In this paper, a new design scheme for actively-shielded x, y-gradient coils, namely, a minimum-power and/or minimum-inductance design scheme using current-loop elements, has been proposed. Its utility in designing MRI gradient coils has been shown by using simulation. An actively-shielded x-gradient coil has been designed as an example and the results are presented. The design scheme seems to be useful for actively-shielded transverse gradient coils, even of non-cylindrical or of arbitrarily -selected shapes.
Keywords