Abstract

In the total artificial heart(TAH), a transcutaneous information transmission system(TITS) is vely important to monitor the TAH status and detect the device failure, and repair the possible problems. First of all, the communication channel(skin) and method were simulated in terms of transmittance, scattering, reflection and absorption, then the system was designed with size reduction including low power consumption and reliability compared to the previous one. The informations are transmitted through the skin(approximately 1cm in depth) by frequency modulated near infrared(NIR) pulses using 780nm laser diodes as transmitters and photodiode as receiver with high speed and high spectral sensitivity. The logic high and low frequencies are 3MHz, 1MHz respectively. The system is a bidirectional data link for more than 38.4Kbps data rate, full-duplex with a bit error rate of less than $10^{-5}$.