고해상도 심전계의 개발

Development of a High-Resolution Electrocardiography

  • 이호성 (건국대학교 의과대학 의학공학과) ;
  • 우응제 (건국대학교 의과대학 의학공학과) ;
  • 박승훈 (건국대학교 의과대학 의학공학과) ;
  • 이종민 (서울대학교 의과대학 의공학과) ;
  • 박광석 (서울대학교 의과대학 의공학과)
  • Lee, H.S. (Department of Biomedical Engineering, College of Medicine, Kon Kuk University) ;
  • Woo, E.J. (Department of Biomedical Engineering, College of Medicine, Kon Kuk University) ;
  • Park, S.H. (Department of Biomedical Engineering, College of Medicine, Kon Kuk University) ;
  • Lee, J.M. (Department of Biomedical Engineering, College of Medicine, Seoul National University) ;
  • Park, K.S. (Department of Biomedical Engineering, College of Medicine, Seoul National University)
  • 발행 : 1996.05.17

초록

Most of the conventional electrocardiogaphies fail to detect signals other than P-QRS-T due to the limited SNR and bandwidth. High-resolution electrocardiography (HRECG) provides better SNR and wider bandwidth for the detection of micro-potentials with higher frequency components such as ventricual late potentials(LP). In this paper, we developed a HRECG using uncorrected XYZ lead. The overall gain of the amplifier is 4000 and the bandwidth is $0.5{\sim}300Hz$ without using 60Hz notch filter. Three 16-bit AH converters sample X, Y, and Z signals simultaneously with a sampling frequency of 2000Hz. Sampled data are transmitted to PC via a DMA-controlled serial communication channel using RS-485 and HDLC protocol. The noise level of the developed HRECG is less than $5{\mu}V_{rms,\;RTI}$. In order to further reduce the noise level, signal averaging technique is implemented utilizing template matching method. The SNR of the developed HRECG is high enough for the detection of LP.

키워드