A Study of Interventricular volume Pressure in a Totally Implantable Electromechanical Artificial heart

완전 이식형 전기기계식 인공심장 심실간 공간 압력 파형에 관한 연구

  • Jo, Yung-Ho (Dept. of Biomedical Engineering, College of Medicine, Seoul National University) ;
  • Choi, Won-Woo (Dept. of Biomedical Engineering, College of Engineering, Seoul National University) ;
  • Park, Seong-Keun (Dept. of Biomedical Engineering, College of Medicine, Seoul National University) ;
  • Lee, Dong-Joon (Dept. of Biomedical Engineering, College of Engineering, Seoul National University) ;
  • Choi, Jae-Soon (Dept. of Biomedical Engineering, College of Engineering, Seoul National University) ;
  • Kim, Hee-Chan (Dept. of Biomedical Engineering, College of Medicine, Seoul National University) ;
  • Min, Byoung-Goo (Dept. of Biomedical Engineering, College of Medicine, Seoul National University)
  • 조영호 (서울대학교 의과대학 의공학 교실) ;
  • 최원우 (서울대학교 대학원 협동과정 의용생체공학) ;
  • 박성근 (서울대학교 의과대학 의공학 교실) ;
  • 이동준 (서울대학교 대학원 협동과정 의용생체공학) ;
  • 최재순 (서울대학교 대학원 협동과정 의용생체공학) ;
  • 김희찬 (서울대학교 의과대학 의공학 교실) ;
  • 민병구 (서울대학교 의과대학 의공학 교실)
  • Published : 1995.11.17

Abstract

인공심장에서의 박출량제어는 전부하에 따른 심박출균형, 심실내 완전충만과 완전구출을 구현하는데 그 목적이 있다. 본 논문에서는 인공심장 내부의 심실간 공간 압력파형을 심박출량의 제어에 활용하였다. 이를 위해서는 심실간 공간의 압력파형을 심실 유입부와 유출부의 혈류량에 대한 정보, 이동작동기의 위치신호와 동기시켜 해석하는 것이 필요하다. 이러한 압력파형의 해석을 바탕으로, 심실간 공간의 음압 크기로 좌,우 심방압을 추정하고 작동기의 좌,우 수축기때 심실간 공간의 음압크기를 비슷하게 유지하도록 작동거리를 제어하면 좌우 심박출 균형을 이룰 수 있음을 확인하였다. 또한, 심실의 완전충만과 완전구출의 실현을 위한 심실 충만 정도를 추정할 수 있게 하는 인자를 제시하였다.

Keywords