Construction of Magnetic Resonance Imaging Inside-out probes for Internal Imaging

핵자기공명 영상법을 위한 내부 영상용 뒤집음-탐침의 제작

  • Ko, R.K. (Dept. of Physics, Paichai Univ.) ;
  • Lee, D.H. (Dept. of Physics, Paichai Univ.) ;
  • Jeong, E.K. (Dept. of Diagnostic Radiology, Yonsei Univ.)
  • 고락길 (배재대학교 물리학과) ;
  • 이동훈 (배재대학교 물리학과) ;
  • 정은기 (연세대학교 의과대학 진단방사선과)
  • Published : 1995.11.17

Abstract

In imaging the samples or human internal organs in a tube shape, general RF-probe types (that encircles a sample or places on top of the sample) are usually unsuitable for the internal imaging due to the degradation of signal-to-noise ratios(SNR's). In the present study the inside-out probes for Magnetic Resonance Imaging (MRI) have been constructed in the three different shapes such as an anti-solenoidal, a saddle and a dual surface types which are positioned as close to the area as possible by putting the probe inside the tubelike sample to improve filling factor. RF-field distributions have also been calculated depending upon the geometrical changes of anti-solenoid probes. Moreover, the performance of the inside-out probes has been checked by measuring SNR's of the images acquired. The inside-out probes constructed in this study produced better SNR's and rf-field uniformity in the area close to the probes in comparing with any other commercial probes. There is a high possibility that the constructed probes in the present study are applicable to the diagnosis of human bodies.

Keywords