Accuracy of Pulsed Doppler Ultrasound Velocity Measurements : In Vitro Flow Phantom Study

Pulsed Doppler 초음파속도측정의 정확도 판정 : 유동 phantom 연구

  • Kim, Young-Ho (Institute of Biomedical Engineering, College of Medicine, Seoul National University) ;
  • Min, Byung-Goo (Institute of Biomedical Engineering, College of Medicine, Seoul National University)
  • 김영호 (서울대학교 의과대학 의공학연구소) ;
  • 민병구 (서울대학교 의과대학 의공학연구소)
  • Published : 1994.12.03

Abstract

An in vitro steady flow experiment was performed in order to test the accuracy of velocity measurement obtained through a pulsed Doppler echocardiography. A flow phantom was designed for the use in a wide velocity range at a given flow rate. The results showed that the pulsed Doppler velocity measurement obtained in this flow phantom is accurate at low flow rates. However, ultrasound velocity measurement should be performed under a careful considerations of PRF and Doppler gain settings, especially at higher flow rates.

Keywords