자기공명 영상촬영을 위한 임의로 선택된 모양의 최소인덕턴스 경사자계코일의 설계

Minimum-Inductance MRI Gradient Coil Design with Arbitrarily-Selected Shape

  • Lee, J.K. (Department of Medical Electronics, Korea University) ;
  • Yang, Y.J. (Department of Medical Electronics, Korea University) ;
  • Yi, Y. (Department of Medical Electronics, Korea University) ;
  • Cho, Z.H. (Department of Information and Communications Engineering, KAIS) ;
  • Oh, C.H. (Department of Medical Electronics, Korea University)
  • 이종권 (고려대학교 의용전자공학과) ;
  • 양윤정 (고려대학교 의용전자공학과) ;
  • 이윤 (고려대학교 의용전자공학과) ;
  • 조장희 (한국과학기술원 (서울) 정보 및 통신공학과) ;
  • 오창현 (고려대학교 의용전자공학과)
  • Published : 1994.05.13

Abstract

This paper proposes a new inductance minimization scheme for a gradient system of arbitrarily selected shape. Although it is important to minimize the gradient coil inductance to reduce the current switching time, such minimization has been possible only for cylindrical or parallel biplanar coils. By using small current loops on arbitrarily selected surface as optimization elements, the inductance of the whole circuit can be minimized using the loop's self- and mutual-inductances. Wire positions can be easily derived from the loop current distribution. Preliminary studies for the design of x-directional surface gradient coil show the utility of tile proposed gradient coil design scheme.

Keywords