A Study on Implemention of Pulse Oximeter System

펄스옥시미터 개발에 관한 연구

  • Park, Y.S. (Dept. of Electronics Eng., Myong Ji Univ.) ;
  • Kim, K.E. (Dept. of Electronics Eng., Myong Ji Univ.) ;
  • Lim, H.S. (Dept. of Electronics Eng., Myong Ji Univ.) ;
  • Yang, D.G. (Dept. of Electronics Eng., Myong Ji Univ.) ;
  • Huh, W. (Dept. of Electronics Eng., Myong Ji Univ.)
  • 박연순 (명지대학교 전자공학과) ;
  • 김근이 (명지대학교 전자공학과) ;
  • 임현수 (명지대학교 전자공학과) ;
  • 양동지 (명지대학교 전자공학과) ;
  • 허웅 (명지대학교 전자공학과)
  • Published : 1993.05.14

Abstract

In this paper, we implemented a pulse wave type oximeter system that continuously measure the of oxyhemoglobin saturation of the arterial blood(SaO2) and pulse rate with non-invasively. We use the transmission type and reflection type transducer for comparison the percentage of hemoglobin oxgen saturation. The light Iron the two kind of LEDs is transmitted through the tissue, then received by a single photodiode with alternatively. By the phase sensitive detection circuit, the output of the photodiode is separated in to red and near infrared signal. We calculated a ratio of light transmittance between two kind of LEDs, and then, it is applied to the oxgen saturation coefficent calculation formular.

Keywords