Fetal heart rate estimation using high resolution pitch detection algorithm

피치 검출 방법을 이용한 태아심음주기의 추출에 관한 연구

  • Published : 1993.05.14

Abstract

Despite the simplicity of processing, conventional autocorrelation function (ACF) method for the precise determination of fetal heart rate (FHR) has many problems. In the case of weak or noise corrupted Doppler ultrasound singnals, the ACF method is very sensitive to the threshold level and data window length. It is real troublesome to extract FHR when there is a data loss. To overcome these problems, the high resolution pitch detection algorithm is adapted to estimate the FHR. The FHR is determined from the correlation of two interconnected segments by its maximum correlation value. FHR is compensated with a constant correlation threshold in a greatly smeared noise signal. This method yields more accurate, robust and reliable than the ACF method.

Keywords