Automatic cardiac output control algorithm for total artificial heart by current waveform analysis

전류파형분석에 의한 완전이식 인공심장의 심박출량 자동제어 알고리즘

  • 최원우 (서울대학교 의용생체공학과) ;
  • 김희찬 (서울대학교 의용생체공학과) ;
  • 민병구 (서울대학교 의용생체공학과)
  • Published : 1993.10.01

Abstract

In this paper, a new automatic cardiac output control algorithm without any pressure sensors for the motor-driven electromechanical total artificial heart(TAH) was developed using motor current information. In the previous studies, many transducers were utilized to obtain informations of hemodynamic states for the automatic cardiac output control. But. such automatic control with sensors has some problems. To solve these problems, I proposed a new "sensorless" automatic cardiac output control algorithm providing the adequate cardiac output to the time-varying physiological demand without right atrial collapse. In-vitro tests were performed to evaluate the performance of a new algorithm and it satisfied the basic three requirements on the pump output response through the mock circulation tests.

Keywords