A Study on Photon Dosimetry System with Diode Defectors

다이오드를 이용한 광자선 선량측정에 관한 연구

  • Lee, D.H. (Dept. of Electronic Eng., Inha University) ;
  • Kang, J.K. (Dept. of Electronic Eng., Inha University) ;
  • Jang, Y.G. (Dept. of Electronic Eng., Inha University) ;
  • Jee, Y.H. (K.C.C.H.) ;
  • Hong, S.H. (Dept. of Electronic Eng., Inha University)
  • Published : 1992.05.15

Abstract

There is a definite requirement to continuously monitor the operating characteristics of radiation therapy machines. It is advisable to monitor the symmetry, flatness, and energy stability of x-ray beams. The semiconductor system was developed using commercially available rectifier diode for th assessment of quality assurance In radiation therapy, which is capable of the above measurements. The beam characteristics of 6MV, 10MV and 21MV photon of Microtron electron accelerator were measured using seven-diodes as detectors and the results were compared with that of using a film results dosimetry with a X-Y plotter. The seven-diode detetor is versatile enough to be used for checking beam profile, flatness, symmetry and energy.