Performance Comparison of the JPED and Full Frame Bit Allocation Techniques for Medical Image Compression

  • Ahn, C.B. (Division of Medical Instrumentations, Korea Academy of Industrial Technology) ;
  • Ro, D.W. (Division of Medical Instrumentations, Korea Academy of Industrial Technology) ;
  • Lee, J.S. (G4 Facsimile Division Korea Academy of Industrial Technology)
  • Published : 1992.05.15

Abstract

The discrete cosine transform (DCT)-based progressive coding standard proposed by the International Standardization Orgnnization (ISO) Joint Photographic Experts Groups (JPEG) is investigated for medical image compression and the performance of the JPEG is compared to that of the full-frame bit-allocation (FFBA) technique. From the comparison, the JPEG standard appeals superior to the FFBA technique in the following aspects: 1) JPEG achieves higher compression than the FFBA technique with less mean square error. 2) Less Gibb's artifact is observed in the compressed images by the JPEG. 3) Computational time for the JPEG is about one-fourth or the FFBA technique. Other attractive points of the JPEG include: Implementation of the JPEG with VLSIs is commercially available in relative low price and the JPEG compression format can easily be interchangeable with other applications.

Keywords